Table 3 Effects of pulse time and pulse delay on nitric oxide uptake efficiency for varied breathing patterns

From: Variability in uptake efficiency for pulsed versus constant concentration delivery of inhaled nitric oxide

  

Uptake efficiency

Pulse

Pulse

Vt = 500 ml

Vt = 500 ml

Vt = 250 ml

Vt = 250 ml

Time [ms]

Delay [ms]

f = 12 min-1

f = 25 min-1

f = 12 min-1

f = 25 min-1

100

0

0.94

0.89

0.88

0.81

 

50

0.94

0.89

0.85

0.72

 

100

0.94

0.89

0.81

0.59

 

200

0.94

0.88

0.70

0.24

200

0

0.94

0.89

0.84

0.70

 

50

0.94

0.89

0.80

0.57

 

100

0.94

0.88

0.76

0.42

 

200

0.94

0.86

0.63

0.14

400

0

0.94

0.88

0.74

0.42

 

50

0.94

0.86

0.68

0.31

 

100

0.94

0.81

0.61

0.22

 

200

0.93

0.62

0.46

0.07